United Orthopedic U-Motion II Cup 10 11 Liner Insertion Instruments F. Impact the Universal Handle until the liner is fully seated. To confirm a complete seating of the liner, palpate around the rim of the cup. XPE & E-XPE Insertion Eliminate any particle on the inner surface of acetabular cup like bone fragment prior to inserting the final liner. It is important to inspect that the cup/liner locking groove is free from debris. Attach a Liner Impactor matched with the selected liner to the tip of the Universal Handle . Place the polyethylene liner into the acetabular cup by hand, making sure that the tabs on the liner are aligned with corresponding tabs in the acetabular cup. There are 12 tabs in the cup that provide variable liner position in 30° increments. Liner Insertion F. Liner Impactor Universal Handle