United Orthopedic U-Motion II Cup 6 7 Cup Insertion Instruments D. Screw Insertion The screw should be inserted in the posterior superior quadrant of the acetabulum on a thick part of the ilium where there is a low-risk area of neurovascular damage. Use the U-Motion II Drill Guide, Drill Bit, and Modular Flex Shaft for drilled hole preparation. The lengths of four Drill Bits are 15, 25, 35 and 50 mm. Carefully drill into acetabulum within selected holes, utilize the Depth Gauge to determine an appropriate screw length. Apical Hole Plugging All U-Motion II acetabular cups can be implanted using the same surgical technique guide. While handling the cup prosthesis, at the proper position and alignment, strike the U-M II Cup Impactor with a mallet until the cup is fully seated. By sighting through the apical hole or screw holes if present, confirm whether a congruent contact between bone- implant interface is achieved. An apical hole cover among the Screw-Hole Cover set is then plugged with a hex head Straight Screwdriver by the Ratchet Handle following cup seating. Supplementary threaded covers are prepared for residual screw holes as well. Cup Insertion Instruments D. Ratchet Handle Straight Screwdriver Surgical tip: Run finger around the edge of dome holes to verify the covers are below the inner surface of the cup. U-M II Drill Guide Modular Flex Shaft Drill Bit Depth Gauge U-M II Cup Impactor