United Orthopedic

United Orthopedic U2 MB Knee Femoral Sizing and Chamfer Resection D. Femoral Sizing and Chamfer Resection D. Instruments AiO Block Threaded Pin 30 mm/50 mm Femoral Stylus Resection Check Blade Remove the Femoral Stylus , then secure the anterior and posterior slots with Threaded Pins and proceed with the anterior and posterior resection. If desired, use the Resection Check Blade to confirm the resection level before bone resection. AiO Block Resection Check Blade Instruments Posterior Referencing: Make sure the bottom knob indicates "0", which refers to the 9 mm standard posterior thickness in the prosthesis. Then remove the Femoral Stylus . Slide the anterior slot to match a proper size on the size panel. Always check the resection level with the Resection Check Blade . “0” 19 20