United Orthopedic
United Orthopedic U2 MB Knee There are two options for preparing tibial platforms. One is the intramedullary alignment method, and the other is the extramedullary alignment. Tibial Intramedullary Alignment Method Access Canal Flex the knee joint to the maximum angle and expose the whole tibial plateau by moving it anteriorly. Use the Step Drill to create an opening into the tibial canal. The drill is inserted to a depth of approximately 100 mm into the tibial canal. After taking out the drill, it is recommended to apply an Alignment Rod into the marrow cavity several times to reduce the risk of fat embolism. Connect the T-Handle to the Tibial IM Rod and insert the assembly manually into tibial canal through the narrowest point inside. Then, remove the T-Handle . If it is difficult to insert or align the Tibial IM Rod , enlarge the pilot hole with the Step Drill again. Proximal Tibial Resection Instruments B. Step Drill Tibial IM Rod Alignment Rod T-Handle Proximal Tibial Resection B. Tibial IM Alignment Guide Tibial Cutting Jig Tibial Stylus Instruments Tibial Cutting Jig Positioning and Tibial Resection Position the Tibial Cutting Jig onto the Tibial IMAlignment Guide . With the thumb screw held loosely, the Tibial Stylus may be used to establish the appropriate height position of the Tibial Cutting Jig . Alignment Rod 5 6
Made with FlippingBook
RkJQdWJsaXNoZXIy MTcyNTM=